ESA Model 584 Solvent Delivery Module
Optimized for HPLC electrochemical detection, the Model 584 Solvent Delivery Module offers pulse-free solvent delivery – even at micro-flow-rate ranges below 50 μl/min – and offers the following features and benefits:
- Dual parallel plungers – provide low pulsation and greater seal life
- Micro-volume plunger displacement – 10 µL per plunger stroke, improves limits of electrochemical detection
- Flow rate range of 0.001 - 10 mL/minute – can be used in analytical, microbore, and LC-MS separations
- Automatic seal wash – prolongs plunger seal life for greater reliability
- Enhancement package – for reduced noise and lower limits of detection with all HPLC detectors
- Electronic diagnostics – for easier maintenance
- Improved fluidics – more reliable and easier to maintain
- High-gradient accuracy and precision – for greater reproducibility
Specifications
| Pump: |
Micro-volume double plunger (approximately 10 µL/stroke) |
| Modes of solvent delivery: |
Constant flow or constant pressure solvent delivery |
| Flow-rate range: |
0.0001 to 10.0000 mL/minute (10 to 340 bar) |
| Flow-rate accuracy: |
No more than 1% or 0.5 µL/minute |
| Flow-rate precision: |
0.3% maximum (RSD: 0.1% maximum) |
| Damper (external): |
Internal capacity of approximately 0.8 mL |
| Materials in contact with mobile phase: |
316 stainless steel, ruby, sapphire, Teflon®, PEEK, Hastelloy® C |
| Time program: |
Commands for flow rate, pressure, event, loop (repeated programs), a total of 10 files and 320 steps |
| Pressure accuracy: |
±10% or 1 mPa, whichever is greater |
| Communications: |
External RS232 converter, optional |
| Size: |
10" (W) x 5" (H) x 16.5" (D) (260 mm x 140 mm x 420 mm), protruding components are not included |
| Weight: |
22 lbs (10 kg) |
| Operating-temperature range: |
41° to 95° F (5° to 35°C) |
| Power requirements: |
90 to 240 VAC, 150 VA , 50/60 Hz |
| Certification: |
USA, Canada, CE, FCC | |