| Sample Capacity: |
Standard tray: 84 vials of 1.8 mL (Std) and 3 vials of 10 mL (LSV) |
| Vial Dimensions (including cap): |
Maximum vial height: 47 mm, Minimum vial height: 32 mm |
| Loop Volume: |
I - 5000 µL (interchangeable)
|
| Dispenser Syringe: |
250 µL (Std) or 1000 µL syringe |
| Vial Detection: |
Missing vial detected by vial sensor |
| Headspace Pressure: |
Built-in compressor |
| Switching Time Injection Valve: |
Electrically <100 msec |
| Piercing Precision Needle: |
± 0.6 mm |
| Wash Solvent: |
External wash solvent bottle |
| Bio-compatible Valve: |
PEEK' injection valve, with quick connect mounting |
| Wetted Parts in Flow Path: |
SS316, PTFE, TEFZEL, VESPEL, Glass, Teflon, PEEK |
| Built-in Column Oven: |
Temperature Range: ambient +5 to 60 C ± II C |
| Oven Dimensions: |
I" W x I" D x 11.811 L (2.5 cm x 2.5 cm x 30 cm) |
| Communication Port: |
Serial RS232C |
| Output: |
Injection marker, 2 Auxiliary outputs, Alarm output (Relay & TTL) |
| Input: |
Next injection, freeze & stop (TR) |
| Ambient Environment: |
10-40'C, 5-95% relative humidity, non-condensing |
| Unit Dimensions: |
30 cm x 50 cm x 34 cm (11.8" H x 19.7" W x 13.4" D) |
| Weight: |
19 kg (42 lb.), 21 kg (46 lb.) with cooling option |
| Power Requirements: |
115/230 VAC; ± 15%; 50/60 Hz; 200 VAC |
| Product Certification: |
CE |
| PERFORMANCE |
| Reproducibility: |
RSD < 0.3% for full loop injections
RSD < 0.5% for partial loop fill injections, injection volumes > 10 µL
RSD < 1.0% for µL pick-up injections, injection volumes > 10 µL
|
| Carry-Over: |
< 0.05% with programmable needle wash |
| PROGRAMMING |
| Methods: |
9 programmable methods with priority sample mode |
| Injection Modes: |
Full loop injection, Partial loop fill injection, µL pick-up injection |
| Injection Volume: |
I µL - 5000 µL, with I µL increments |
| Injections per Vial: |
Maximum 9 injections |
| Analysis Time: |
Maximum 9 hr. 59 min. 59 sec. |
| Programmable Needle Wash: |
Wash between injections, between vials, between series |
| Programmable Timed Events: |
4 x ALJX-1 on/off, 4 x AL7X-2 on/off, Initial oven setpoint, 2 x new oven setpoint |
| Series: |
9 programmable series |
| SAMPLE TRAY COOLING |
| Temperature Control: |
Integrated, Peltier type |
| Temperature Range/Capacity: |
4-15º C, no greater than 20º' C below ambient |
| Temperature Accuracy: |
± 2º C |