Analytical Thermostatted Autosampler for Electrochemical Detection
The Analytical Thermostatted Autosampler is designed for electrochemical detection with the Thermo Scientific Dionex highly sensitive Coulochem III electrochemical detector.
This autosampler is ideally suited for use with electrochemical detection when using the highly sensitive Coulochem III electrochemical detector. The in-line split-loop (or flow-through) injection design (needle and sample loop integral to the high-pressure fluidic path) and external needle wash option virtually eliminate carryover. A clear advantage of this design is the compatibility with smallest injection volumes and zero sample loss during the injection procedure. In addition, the sampler is designed for best precision, accuracy, and speed. The autosampler is highly flexible with a wide range of supported sample formats, customizable pre-column derivatization, and integrated sample cooling. Features include:
- Supports various sample formats including different sized vials, well plates, and Eppendorf tubes
- User-defined programs for precolumn derivatization of neuroactive amino acids and other compounds
- Injection volume precision typically <0.15% RSD for 5 μL injections
- Superior injection-volume linearity of r²=0.999 (5–90 μL) due to a high-precision drive mechanism
- Superior volumetric accuracy of better than ±0.5% for consistent results
- Sample thermostatting for optimal protection of thermally sensitive analytes
- Supports an injection volume range from 1–100 μL
Specifications
|
Sample capacity
|
216 × 0.3 mL vials,
120 × 1.1 mL conical vials,
216 × 1.2 mL vials,
120 × 1.8 mL/2.0 mL vials,
66 × 4 mL vials,
30 × 10 mL vials,
3 x 24 deep well plates, 96 and/or 384 normal or deep well plates,
3 x 40 0.5 mL and/or 1.5 mL Eppendorf tubes, and/or 3 x 384 low well PCR plates
+ 15 × 10 mL vials (depending on sample tray configuration)
|
|
Injection methods
|
In-line split-loop (flow-through), bypass mode, user-defined programs
|
|
Injection volume range (recommended)
|
0.01–100 µL (1–100 µL)
|
|
Injection volume accuracy
|
±0.5% at 50 and 90 µL
|
|
Minimum sample required
|
1 μL out of 5 μL (250 μL conical vial)
|
|
Injection volume precision
|
< 0.25% RSD at 5 μL (typically < 0.15% RSD),
caffeine in water
|
|
Linearity
|
Corr. coeff. >0.9999, RSD <0.5% at 5–90 μL, caffeine in water
|
|
Maximum operating pressure
|
<345 Mpa (5,000 psi)
|
|
Needle wash
|
Active external needle wash
|
|
Carry over
|
< 0.004% for caffeine at 7.5 MPa with external wash
|
|
Injection cycle time
|
<15 s for 5 µL
|
|
Sample thermostatting
|
4–45 °C or 22 °C below ambient
|
|
Sample temperature accuracy
|
±2 °C
|
|
Dimensions
|
36 × 42 × 51 cm (14.2 x 16.5 x 20 in.)
|
|
Power requirements
|
100–120 V, 60 Hz
200–240 V, 50 Hz
|
|
PC connection
|
All functions controllable via USB;
integrated USB hub with 3 USB 1.1 ports
|
|
Weight
|
max. 23 kg (51 lbs) including cooling
|
|
I/O interfaces
|
4 digital inputs, 4 relay outputs
|
* Specifications subject to change without notice |